Display
8-inch color touch Screen Liquid Crystal Display (LCD)
Parameter
WBC, Neu#, Lym# , Mid# , Neu%, Lym%, Mid%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCR
Histograms
WBC, RBC, PLT histogram
Calibration
Manual and auto calibration
Quality control
3 Level QC, LJ graph
Sample Volume
Venous mode : 10L Venous blood
Capilary mode: 10L capilary blood
Predilute Mode: 20L Capilary blood
Resolution
800 600
Storage
100,000 Sample results with histograms
Reagent
Lyse (500ml) and Diluent (20L)
Printout
Thermal printer, support external printer
Temperature
18 - 35°C
Power Supply
AC 100-240V, 50/60 ± 1Hz
Dimension
410mm x 435mm 472mm
MEASUREMENT RANGE
WBC: (0.99.9 x 109) VL, RBC: (0-99.9 x 1012)VL, HGB: (0-300.0)L, PLT(0-999 x 109)L
PERFORMANCE
PRECISION (CV)
WBC =2.0% (7.O – 15.0 x 109) L RBC =1.9% (3.5 – 6.0 x 1012) L
HGB =1.9% (110.0 – 180.0g) L MCV =0.4% (80.0 – 110.0) Fl
LT =4.0% (100.0 – 500.0 x 109)vL